Asia Mattelab

Flame Photometer

Flame Photometer

A.KRÜSS flame photometer for the analysis of alkali and alkaline earth elements

Economical model for the laboratory with very high precision.
Measurement range FP8400:
Na: 0.01-4500ppm; 0.0004-200mmol/l;
K: 0.01-4500ppm; 0.0003-110mmol/l;
Li: 0.01-4500ppm; 0.0014-600mmol/l;
Ca: 0.50-4500ppm; 0.0125-110mmol/l

Discription

The A.KRÜSS flame photometers are used to determine the concentration of alkali and alkaline earth elements in aqueous solutions. The devices are a simple and particularly cost-effective alternative to analysis techniques such as ICP or AAS.

Flame photometers are particularly suitable for the quantitative determination of alkali metals such as sodium (Na), potassium (K) and lithium (Li) and alkaline earth metals such as calcium (Ca). These elements are measured simultaneously in a single work step.

Consistent modernization of the flame photometers and a special focus on safety concepts open up completely new fields of application for many cost-sensitive customer groups – from automated incoming goods inspection to the qualification of FFP masks in accordance with DIN 149.
Low operating costs are one of the main strengths of AES, and the technology of the FP8000 series also scores with high accuracy.

Technical data

TABLE
DISPLAYTFT display with integrated 8.4” TFT touchscreen, 800 x 600 pixels
INTERFACES2 x USB
1 x Ethernet
1 x RS-232 for printer
2 x USB
1 x Ethernet
1 x RS-232 for printer
Upgrade 4-20 mA analog, passive current interface
2 x USB
1 x Ethernet
1 x RS-232 for printer
2 x USB
1 x Ethernet
1 x RS-232 for printer
24-character plain paper dot matrix printer
ACCESSORIESPressure reducing station with filter units
Various standards and system solutions
Spare parts and accessories direct from the manufacturer
RETROFITTABILITYRetrofit kits available for other models
FP8400FP8500FP8600FP8700
CALIBRATIONLinear with 2 standards
Non-linear with 6-8 standards, polynomial curve fit
DRIFT1 % in 60 min
REFERENCE LINE (optional)Lithium guideline (reference line) with 5 mmol/l
Caesium guideline on request
SAMPLE VOLUME2.5 ml2.5 ml2.5 mlmicroliter range (depending on dilution ratio and sample tube)
COMBUSTION GASPropane: recommended for alkaline elements
Butane: possible propane substitute

Model specifications

TABLE